PX-VT001 Portable Ventilator for Emergency Medical Situations
ENVIRONMENTAL REQUIREMENTS
- Normal working conditions:Temperature:-18~50℃
- Relative Humidity:≤95%(No condensation)
- Atmospheric Pressure:70kPa~ 110kPa
- Gas source:medical oxygen of 280~ 600kPa
- Power:AC 100-240V,50/60Hz,30VA
- Internal electrical power source:Li-lon Battery 12V/2200mAh
CLASSIFICATION
- Class l
- Type B applied part
- IP21
- Transportable
STANDARD CONFIGURATION
- Main unit,
- Mask,Active exhalation valve,Respiratory pipeline,
- Exhalation control sampling tube,Oxygen tube,Oxygen regulator
- (Different standard pressure reducing valvescan be equipped according to user needs).
OPTIONS
- Oxygen cylinder
MAIN PARAMETERS
- Respiratory control mode:CMV
- Respiratory rate:5bpm-40bpm
- Ventilation flow rate: 0->20L/min
- I/E: 1:2
- FiO2: Air Mix or No Air Mix
- Battery life:>8hours
SAFETY MEASURES
- Low voltage:Medium priority alarm
- Battery depletion:High priority alarm
- High airway pressure:Issue a medium priority alarm when the airwaypressure is higher than 5.5kPa;
- Automatic pressure relief protection whenairway pressure exceeds 8kPa